The Quality of Dialysis Water: A Case Study in the Educational Hospitals of Yazd, Iran

Mohammad Saleh Ali Taleshi 1 , * , Farhad Nejadkoorki 1 , Hamid Reza Azimzadeh 1 , Seyedeh Mahdieh Namayandeh 2 , Motahhareh Sadat Namayandeh 1 and Mohammad Taghi Ghaneian 3
Authors Information
1 Department of Environment, Faculty of Natural Resources, Yazd University, Yazd, IR Iran
2 Cardiovascular Research Center, Shahid Sadoughi University of Medical Sciences, Yazd, IR Iran
3 Department of Environmental Health, Shahid Sadoughi University of Medical Sciences, Yazd, IR Iran
Article information
  • Health Scope: February 19, 2015, 4 (1); e23086
  • Published Online: February 10, 2015
  • Article Type: Research Article
  • Received: August 26, 2014
  • Revised: October 30, 2014
  • Accepted: November 19, 2014
  • DOI: 10.17795/jhealthscope-23086

To Cite: Taleshi M S A, Nejadkoorki F, Azimzadeh H R, Namayandeh S M, Namayandeh M S, et al. The Quality of Dialysis Water: A Case Study in the Educational Hospitals of Yazd, Iran, Health Scope. 2015 ;4(1):e23086. doi: 10.17795/jhealthscope-23086.

Abstract
Copyright: Copyright © 0, Journal of Health Scope. This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://creativecommons.org/licenses/by-nc/4.0/) which permits copy and redistribute the material just in noncommercial usages, provided the original work is properly cited.
1. Background
2. Objectives
3. Materials and Methods
4. Results
5. Discussion
Acknowledgements
Footnote
References
  • 1. Arvanigidou M. Microbiological quality of water and dialysate in all haemodialysis centres of Greece. Nephrol Dial Transpl. 1998; 13(4): 949-54[DOI]
  • 2. Hoenich NA, Levin R. The implications of water quality in hemodialysis. Semin Dial. 2003; 16(6): 492-7[PubMed]
  • 3. Pontoriero G, Pozzoni P, Andrulli S, Locatelli F. The quality of dialysis water. Nephrol Dial Transplant. 2003; 18 Suppl 7-5[PubMed]
  • 4. de Torres JP, Strom JA, Jaber BL, Hendra KP. Hemodialysis-associated methemoglobinemia in acute renal failure. Am J Kidney Dis. 2002; 39(6): 1307-9[PubMed]
  • 5. Rao TK, Friedman EA. Editorial: Fluoride and bone disease in uremia. Kidney Int. 1975; 7(3): 125-9[PubMed]
  • 6. Manzler AD, Schreiner AW. Copper-induced acute hemolytic anemia. A new complication of hemodialysis. Ann Intern Med. 1970; 73(3): 409-12[PubMed]
  • 7. D'Haese PC, De Broe ME. Adequacy of dialysis: trace elements in dialysis fluids. Nephrol Dial Transplant. 1996; 11 Suppl 2: 92-7[PubMed]
  • 8. BSI Group Headquarters , inventor . Draft BS ISO 13959 Water for Hemodialysis and Related Therapies. 2013;
  • 9. Kawanishi H, Masakane I, Tomo T. The new standard of fluids for hemodialysis in Japan. Blood Purif. 2009; 27 Suppl 1: 5-10[DOI][PubMed]
  • 10. Vorbeck-Meister I, Sommer R, Vorbeck F, Horl WH. Quality of water used for haemodialysis: bacteriological and chemical parameters. Nephrol Dial Transplant. 1999; 14(3): 666-75[PubMed]
  • 11. Asadi M, Norouzi M, Khazaei M, Omidi Oskouei A, Paydari Shayesteh N. [Concentrations of Cations in Water Used for Hemodialysis in Qom Province Hospitals and Comparisons with AAMI and EPh Standards]. J. Health hygiene. 2011; 2(3): 48-55
  • 12. Sobrino Perez PE, Barril Cuadrado G, del Rey Roman C, Sanchez Tomero JA. [Monitoring on-line treated water and dialysate quality]. Nefrologia. 2008; 28(5): 493-504[PubMed]
  • 13. Laurence RA, Lapierre ST. Quality of hemodialysis water: a 7-year multicenter study. Am J Kidney Dis. 1995; 25(5): 738-50[PubMed]
  • 14. Alizadeh M, Bazrafshan E, Jafari Mansoorian H. Microbiological and Chemical Indicators of Water Used in Hemodialysis Centers of Hospitals Affiliated to Zahedan University of Medial Sciences. JHAD. 2012; 2(3): 182-91
  • 15. Lenore SC. Standard Methods for the Examination of Water and Wastewater. 1999;
  • 16. Ibrahim S. Quality of care assessment and adherence to the international guidelines considering dialysis, water treatment, and protection against transmission of infections in university hospital-based dialysis units in Cairo, Egypt. Hemodial Int. 2010; 14(1): 61-7[DOI][PubMed]
  • 17. Tonelli M, Wiebe N, Hemmelgarn B, Klarenbach S, Field C, Manns B, et al. Trace elements in hemodialysis patients: a systematic review and meta-analysis. BMC Med. 2009; 7: 25[DOI][PubMed]
  • 18. Baseri A, Dehghani R, Soleimani A, Hasanbeigi O, Pourgholi M, Ahaki A, et al. Water Quality Investigation of the Hemodialysis Instruments in Kashan Akhavan Hospital During Oct.-Nov. 2011. Iran J Health Env. 2013; 6(2): 145-54
  • 19. Shahryari A, Nikaeen M, Hajiannejad M, Vahid Dastjerdi M, Hassanzadeh A, Saffari H, et al. Evaluation of Reverse Osmosis System for Improving Water Quality Used for Dialysis Fluid. Heelth Serv Res J. 2012; 8(5): 782-95
  • 20. Asadzadeh SN, Nemati Sani O, Sajjadi S.A. , Yousefi Sangani M. . Chemical Quality of Water Entrance to Dialysis Machines and Its Comparison with AAMI and EPh Standards in Hospital of 22 Bahman gonabad. JNKUMS. 5: 1137-43
Creative Commons License Except where otherwise noted, this work is licensed under Creative Commons Attribution Non Commercial 4.0 International License .

Search Relations:

Author(s):

Article(s):

Create Citiation Alert
via Google Reader